Agilent Technologies E8257D/67D Portable Generator User Manual


 
Chapter 3 23
Operation Verification
Checking the Maximum Leveled Power
3. Ensure that the RF OUTPUT connector is connected to a 50 load.
4. Ensure that the power level entered corresponds to the value listed in Table 3-1.
5. Refer to Table 3-2 for the recommended equipment and measure the output power.
If the measured power level is more than the power level listed in Table 3-1, turn the front panel knob
counterclockwise until the measured power level equals the power level in Table 3-1.
6. Note the signal generator front-panel amplitude value and use it to repeat the maximum leveled power
check.
If all of the above steps are followed and the check continues to fail, contact the nearest Agilent
Technologies call center for assistance. Refer to “Contacting Agilent Sales and Service Offices” on
page 16 for contact information.
Table 3-1 Frequency and Power Level Limits
Instrument Option Amplitude
Standard Option 1EA
Start
Frequency
Stop
Frequency
# Points
E8257D
Option 520
13 dBm 16 dBm 250 kHz 3.2 GHz 50
13 dBm 20 dBm 3.21 GHz 20 GHz 200
E8257D
Option 540
9 dBm 15 dBm 250 kHz 3.2 GHz 50
9 dBm 18 dBm 3.21 GHz 20 GHz 200
9 dBm 14 dBm 20.1 GHz 40 GHz 200
E8257D
Option 550 and
Option 567
5 dBm 14 dBm 250 kHz 3.2 GHz 50
5 dBm 16 dBm 3.21 GHz 20 GHz 200
5 dBm 14 dBm 20.1 kHz 60 GHz 200
5 dBm 12 dBm 60.1 GHz 67 GHz 100
E8267D
Option 520
13 dBm N/A 250 kHz 3.2 GHz 50
18 dBm N/A 3.21 GHz 20 GHz 200
E8267D
Option 532 and
Option 544
12 dBm N/A 250 kHz 3.2 GHz 50
14 dBm N/A 3.21 kHz 20 GHz 200
12 dBm N/A 20.1 GHz 40 GHz 200
10 dBm N/A 40.1 GHz 44 GHz 50