3.1 Setting the Parameters
3-5
Example:
Suppose that the following parameters are set for this unit:
Output Level :
10 dBm
Digital Mod : On
System : Noise1
Noise Bandwidth : 7.68 MHz
Calculated Bandwidth : 3.84 MHz
From Formula (1),
Power within the calculated bandwidth
=
RF output level
+
Calculated level
=
10 dBm
+
(
3.45 dB)
=
13.45 dBm
Therefore, the power within the calculated
bandwidth for the RF output is –13.45 dBm.
The power –13.45 dBm is displayed at “(Ab-
solute)”.