7.2.1 Advanced Configuration Menu Settings
The Advanced Configuration menu offers settings to customize the Atlas Monitor to suit your needs. Not
all options are available on all models.
Language - Factory default will be set for the
dominant language used in your country. Of
course, if you are more familiar with another
language, the monitor provides a choice of eight.
Silence duration - The large
Silence button on
the right edge of the monitor allows you to turn
off all of the alarms simultaneously – for a short
duration lasting either 60, 90, or 120 seconds.
ECG gain - When 10 mm/mV gain is chosen,
the height of the vertical ruler that appears to the
left of the ECG waveform is 10 mm high and
indicates a 1 mV amplitude. When automatic
gain is selected the ruler height will vary, but it
will always indicate a 1 mV signal size. The ruler
size is automatically increased or decreased to
scale to a particular set of waves, but the vertical
line still indicates the same amplitude of 1 mV.
ECG lead set - The Atlas Monitor allows for
both 3 wire and 5 wire ECG lead sets.
ECG speed - The amount of ECG waveform
shown on the CRT is determined by the trace
speed. A slower trace speed means more
seconds of waveform are shown on the CRT.
ECG bandwidth - The ECG waveform can be
displayed and printed in either Monitor or
Extended bandwidth. Monitor mode allows for
a clearer picture of the waveform by filtering out
noise. Extended mode, usually used with
cardiac paced patients, shows the finer nuances
of ECG waveform facilitating the detection of
conditions such as ischemia.
Initial pressure – Specifies the initial blood
pressure cuff pump-up pressure. If this pressure
is not enough to complete a blood pressure
measurement, the cuff will re-inflate with a
higher target pressure and re-attempt the
measurement.
MAP - The Mean Arterial Pressure can be
displayed on the screen, above the ECG
waveform. This measurement is calculated from
the Systolic and Diastolic readings. The choices
are Yes (MAP is displayed) or No (MAP is not
displayed).
Alternate BP field (when language setting is
configured to Chinese only) – You can display
MAP or blood pressure in kPa on the screen
(upper left) above the ECG waveform. The
choices are MAP, kPa, or nothing. If you
choose “nothing” then, neither MAP nor kPa
will be displayed on the screen.
Second trace selection – All models offer a
choice of displaying a second line of ECG
waveform or displaying the SpO
2
waveform.
Model 622xx adds the option of displaying
Impedance Respiration as the second trace.
Model 623xx adds the option of displaying CO
2
concentration as the second trace.
Temperature units - In models 622xx and
623xx, temperature can be displayed as either
Celsius °C, or Fahrenheit °F.
Respiration speed – Models 622xx and 623xx
allow the Impedance Respiration or CO
2
waveforms to be displayed at three different
speeds.