Welch Allyn Medical Diagnostic Equipment 621NO Blower User Manual


 
3
Trend Data - Trend data is captured every time
blood pressure is measured, whether this event is
automatic or manual. Trend data is also
automatically captured every 15 minutes if blood
pressure intervals are longer, or blood pressure is
not used. Push the
Trend button to see the trend
data. Scroll through the trend data with either
Set
button. Push
Trend again to return to the
waveform display. The monitor will hold up to
144 lines of trend data, which is 36 hours if data is
captured every 15 minutes.
Printing - A printer option is available with both
models 621xx and 622xx, and is a standard model
623xx feature. Push the
Print button to print what
is on the screen. If the waveforms are displayed
on the screen, the
Print button prints 15 seconds
of waveforms plus all the current measurements.
The printout captures data from 9 seconds before
the
Print button was pressed until 6 seconds after.
Note: If any measurements exceed the alarm
limits, they will be marked with asterisks on
the printout. If trend data is displayed on the
screen,
Print prints all the trend data. If the
Print On Alarm configuration is set to Yes, and
a Patient Alarm or Measurement Invalid Alarm
occurs, Atlas will print the current
measurements and the currently configured
waveforms automatically.
If your model of the Atlas Monitor does not have
the optional printer, the
Print button is labeled
Freeze, and it freezes, or halts, the waveform
display for 10 seconds to permit studying of the
waveform.