Welch Allyn Medical Diagnostic Equipment 621NO Blower User Manual


 
18
ECG settings in Advanced Configuration -
The five settings in the Advanced Configuration
menu associated with the ECG are listed in this
table:
Advanced Configuration
ECG setting
Possible values
ECG gain automatic |
10 mm/mV
ECG lead set 3 wire | 5 wire
ECG speed 6.25 | 12.5 | 25 mm/s
ECG bandwidth Monitor | Extended
Second trace
selection
ECG | SpO
2
|
Respiration | CO
2
You can change the ECG settings in Advanced
Configuration:
Press the Date/Time button.
Press
Trend.
Use either
Select button to highlight ECG
gain, ECG lead set , ECG speed, ECG
bandwidth, or Second trace selection.
Use the
Set button to change any of these
values.
Press
Trend to exit Advanced Configuration
ECG gain - The height of the vertical ruler that
appears to the left of the ECG waveform
indicates a 1 mV amplitude and is 10 mm high if
10 mm/mV gain is chosen. When automatic
gain is selected the ruler height will vary, but it
will always indicate a 1 mV signal size. The
ruler size is automatically increased or decreased
to scale to a particular set of waves, but the
vertical line still indicates the same amplitude of
1 mV.
ECG lead set - The Atlas Monitor allows for
both 3 wire and 5 wire ECG lead sets.
Note: When using 3 wire leads, the ECG lead
set must be set correctly. Incorrect results
and noisy waveforms can be obtained if the
system is configured for 5 wire leads when
using 3 wire leads.
ECG speed - The amount of ECG waveform
shown on the CRT is determined by the trace
speed. A slower trace speed means more
seconds of waveform are shown on the CRT.
ECG bandwidth - The ECG waveform can be
displayed and printed in either Monitor or
Extended bandwidth. Monitor mode allows for
a clearer picture of the waveform by filtering out
noise. Extended mode, usually used with
cardiac paced patients, shows the finer nuances
of ECG waveform, facilitating the detection of
conditions such as ischemia.
Note: Detection of ischemia is the interpretation
of the clinician only, the Atlas Monitor does
not provide automated ischemia detection.
Note: It is normal for the ECG baseline to
wander slightly in Extended bandwidth.
Pacemaker signals – The Atlas Monitor
displays pacemaker signals exactly as they are
captured. There is no option to display symbolic
pace indications. Use Extended bandwidth for
enhanced display of pacemaker signals.