23
3.2 Laser plummet
[Laser plummet model]
The laser plummet is not set to be ON at factory shipping.The laser plummet operation of
power supply ON can be set by command No 520, LD PLUM & E VIAL. For using Command
number, refer to 9-2. Accessing by 007.
[For the Detaching type laser plummet equipment model]
Turn on the laser plummet function by pushing the Laser key.
Match the position with the leveling screw so that the laser mark coincides with the
ground mark.
[For the Shift type laser plummet equipment model]
• Turn on the laser plummet function by pushing the Laser key.
• Match the position by the tripod so that the laser mark coincides with the ground mark.
• The centering knob is loosened, and the upper plate is pushed by the tip of a finger,
and a center mark is matched to the ground mark.
• Tighten the centering knob.
• Loosen the horizontal clamp screw, and rotate the instrument by 90°,and confirm the
vial of the circular vial is at the center at any position.
Correct the vial with the leveling screw when the vial comes off from the center.
[Brightness adjustment of laser]
Sometimes the state of the surface of the ground mark or a surrounding environmental
dose not allow observing the laser spot easily. Please adjust the brightness of the laser if
necessary.
If the Laser key is pressed, the brightness adjustment
screen of the laser plummet device, is displayed.
The Laser plummet becomes dark by,
key and becomes bright by / key
The adjustment is completed with the ENT key and it
returns to electronic vial screen.
LD PLUM.POWER ADJ
DOWN
0
UP
10
5
ENT
ELECTRONIC VIAL
T.COMP.
ON
30”/1DIV
SENS.
LD PLUM.POWER ADJ
DOWN
0
UP
10
5
LD POINT PLUM ADJ
ENT