Anritsu MG3690A Portable Generator User Manual


 
7
T
ypical maximum MG3694A available output power
0 to 25 dB or to minimum rated power
, whichever is higher
T
ypical
T
ypical maximum MG3692A available output power
Other Output Power Specifications
Output Units: Output units selectable as either dBm or mV.
Selection of mV assumes 50
load. All data entry and display
are in the selected units.
Output Power Resolution: 0.01 dB or 0.001 mV
Source Impedance: 50nominal
Source SWR (Internal Leveling): <2.0 typical
Power Level Stability with Temperature: 0.04 dB/deg C typical
Level Offset: Offsets the displayed power level to establish a new
reference level.
Output On/Off: Toggles the RF output between an Off and On
state. During the Off state, the RF oscillator is turned off. The On
or Off state is indicated by two LEDs located below the OUTPUT
ON/OFF key on the front panel.
RF On/Off Between Frequency Steps: System menu selection of
RF On or RF Off during frequency switching in CW, Step Sweep,
and List Sweep modes.
RF On/Off During Retrace: System menu selection of RF On or RF
Off during retrace.
Internal Leveling: Power is leveled at the output connector in
all modes.
External Leveling:
External Detector: Levels output power at a remote detector
location. Accepts a positive or negative 0.5 mV to 500 mV input
signal from the remote detector. EXT ALC ADJ adjusts the input
signal range to an optimum value. BNC connector, rear panel.
External Power Meter: Levels output power at a remote power
meter location. Accepts a ±1V full scale input signal from the
remote power meter. EXT ALC ADJ adjusts the input signal
range to an optimum value. BNC connector, rear panel.
External Leveling Bandwidth: 30 kHz typical in Detector mode.
0.7 Hz typical in Power Meter mode.
User Level Flatness Correction:
Number of points: 2 to 801 points per table
Number of tables: 5 available
Entry modes: GPIB power meter or computed data
CW Power Sweep
Range: Sweeps between any two power levels at a single CW
frequency.
Resolution: 0.01 dB/step (Log) or 0.001 mV (Linear)
Accuracy: Same as CW power accuracy.
Log/Linear Sweep: Power sweep selectable as either log or linear.
Log sweep is in dB; linear sweep is in mV.
Step Size: User-controlled, 0.01 dB (Log) or 0.001 mV (Linear) to
the full power range of the instrument.
Step Dwell Time: Variable from 1 ms to 99 seconds. If the sweep
crosses a step attenuator setting, there will be a sweep dwell of
approximately 20 ms to allow setting of the step attenuator.
Sweep Frequency/Step Power
A power level step occurs after each frequency sweep. Power level
remains constant for the length of time required to complete each
sweep.
Internal Power Monitor (Option 8)
Sensors: Compatible with Anritsu 560-7, 5400-71, or 6400-71
series detectors. Rear panel input.
Range: +16 dBm to –35 dBm
Accuracy: ±1 dBm, (+16 to –10 dBm)
±2 dBm, (–10 to –35 dBm)
Resolution: 0.1
dBm minimum
Accuracy and Flatness
Accuracy specifies the total worst case accuracy. Flatness is
included within the accuracy specification.
Analog Sweep Mode (typical):
Step Sweep and CW Modes:
Accuracy:
0-25 dB
±1.0 dB ±1.5 dB ±1.5 dB ±1.5 dB
25-60 dB
±1.0 dB ±1.5 dB ±3.5 dB
N/A
60-100 dB
±1.0 dB ±1.5 dB
±
3.5 dB
N/A
Flatness:
0-25 dB
±0.8 dB ±1.1 dB ±1.1 dB ±1.1 dB
25-60 dB
±0.8 dB ±1.1 dB ±3.1 dB
N/A
60-100 dB
±0.8 dB ±2.1 dB
±
3.1 dB
N/A
Accuracy:
0-12 dB ±2.0 dB ±2.0 dB ±2.0 dB ±3.0 dB
12-30 dB
±3.5 dB ±3.5 dB ±4.6 dB ±5.6 dB
30-60 dB
±4.0 dB ±4.0 dB ±5.2 dB ±6.2 dB
60-122 dB ±5.0 dB ±5.0 dB ±6.2 dB ±7.2 dB
Flatness:
0-12 dB ±2.0 dB ±2.0 dB ±2.0 dB ±2.5 dB
12-30 dB ±3.5 dB ±3.5 dB ±4.1 dB ±5.1 dB
30-60 dB ±4.0 dB ±4.0 dB ±4.6 dB ±5.6 dB
60-122 dB ±5.0 dB ±5.0 dB ±5.2 dB ±6.2 dB
Attenuation Frequency (GHz)
Below
Max Power
40 40-50 50-60 60-65
Attenuation Frequency (GHz)
Below
Max Power
0.01-0.05 0.05-20 20-40 40-65