Agilent Technologies 54753a Sprinkler User Manual


 
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4}:RISetime[?]
<risetime>
Sets the risetime for the normalized response. The risetime setting is limited
by the timebase settings and the record length.
<risetime>
Risetime setting in seconds.
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4}:TDRDest[?]
CHANnel<number>
Selects a TDR destination channel for an external stimulus. When you use an
external stimulus, you must use this command to specify where the TDR
channel is coming into the instrument.
<number>
An integer, 1 through 4, indicating the slot in which the channel resides,
followed by an optional A or B identifying which of two possible channels in
the slot is being referenced.
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4}:TDRTDT[?]
{TDR | TDT}
Sets the instrument to make either TDR or TDT measurements. When the
setting is TDT, both TDR and TDT measurements can be made
simultaneously. You must select a TDT destination channel before selecting
TDT.
Command
:TDR{2 | 4}:RESPonse{1 | 2 | 3 | 4}:TDTDest[?]
{NONE | CHANnel<number>}
Selects a destination channel for a normalization measurement.
<number>
An integer, 1 through 4, indicating the slot in which the channel resides,
followed by an optional A or B identifying which of two possible channels in
the slot is being referenced.
TDR Commands
39