Agilent Technologies 54753a Sprinkler User Manual


 
Command
:MEASure:RISetime[?] [<source>]
Measures the rise time of the first displayed edge by measuring the time at
the lower threshold of the rising edge, measuring the time at the upper
threshold of the rising edge, then calculating the rise time.
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number>}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Command
:MEASure:SOURce[?] <source>[,<source>]
Selects the source for measurements. Two sources can be specified with this
command. All measurements except MEASure:DELTatime are made on the
first specified source. The delta time measurement uses two sources if two
are specified. If only one source is specified, the delta time measurement
uses that source for both of its parameters.
<source>
{CHANnel<number> | FUNCtion<number> |
RESPonse<number> | WMEMory<number> | HISTogram |
FFT}
<number>
For channels: an integer, 1 through 4, indicating the slot in which the channel
resides, followed by an optional A or B identifying which of two possible
channels in the slot is being referenced. For functions: 1 or 2. For waveform
memories (WMEMory): 1, 2, 3, or 4. For TDR responses: 1, 2, 3, or 4.
Measure Commands
29