Agilent Technologies 54622A Sprinkler User Manual


 
3-11
Testing Performance
To verify voltage measurement accuracy
b Set the Volts/Div setting to the value in the first line in Table 3-5.
Table 3-5
Settings Used to Verify Voltage Measurement Accuracy
c Press the Acquire key. Then press the Averaging softkey and set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2
Press the Cursors key, set the Mode softkey to Normal, then press the XY
softkey and select
Y. Press the Y1 softkey, then use the Entry knob
(labeled
on the front panel) to set the Y1 cursor on the baseline of
the signal.
Volts/Div Setting Power Supply Setting Test Limits
5 V/Div 35 V 34.04 V to 35.96 V
2 V/Div 14 V 13.616 V to 14.384 V
1 V/Div 7 V 6.808 V to 7.192 V
0.5 V/Div 3.5 V 3.404 V to 3.596 V
0.2 V/Div 1.4 V 1.3616 V to 1.4384 V
0.1 V/Div 700 mV 680.8 mV to 719.2 mV
50 mV/Div 350 mV 340.4 mV to 359.6 mV
20 mV/Div 140 mV 136.16 mV to 143.84 mV
10 mV/Div 70 mV 68.08 mV to 71.92 mV
5 mV/Div 35 mV 34.04 mV to 35.96 mV
2 mV/Div 14 mV 13.616 mV to 14.384 mV
1 mV/Div* 7 mV 6.616 mV to 7.384 mV
*Full scale is defined as 16 mV on the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.