ADDITIONAL MEASUREMENT PROCEDURE
For samples containing chlorine
Select the program number corresponding to Ozone on the secondary
LCD by pressing PROGRAM and .
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP”
will blink on the
display.
Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Remove the cuvet.
Add the content of one
packet of the optional
reagent HI 93703-52-0
Glycine Powder. Replace the
cap and shake gently until
dissolution is complete.
Add the content of one
packet of HI 93757-0
Ozone Reagent. Replace
the cap and shake gently
for 20 seconds.
Replace the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
10 mL
Ozone