10 mL
SPECIFICATIONS
Range 0.00 to 5.00 mg/L
Resolution 0.01 mg/L from 0.00 to 2.50 mg/L;
0.10 mg/L above 2.50 mg/L
Precision ±0.03 mg/L @ 1.00 mg/L
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference filter
@ 525 nm
Method Adaptation of the USEPA method 330.5 and
Standard Methods for the Examination of Water
and Wastewater, 20
th
edition, 4500-Cl G. The
reaction between free chlorine and the DPD reagent
causes a pink tint in the sample.
REQUIRED REAGENTS
Code Description Quantity/test
HI 93701-0 DPD Powder Reagent 1 packet
REAGENT SETS
HI 93701-01 Reagents for 100 tests
HI 93701-03 Reagents for 300 tests
For other accessories see page 49.
MEASUREMENT PROCEDURE
Select the program number corresponding to Free Chlorine on the
secondary LCD by pressing PROGRAM and .
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
FREE CHLORINE
Free Chlorine