Ozone
STANDARD MEASUREMENT PROCEDURE
Select the program number corresponding to Ozone on the secondary
LCD by pressing PROGRAM and .
Fill the cuvet with 10 mL of unreacted
sample, up to the mark, and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP”
will blink on the
display.
Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Remove the cuvet.
Add the content of one packet
of HI 93757-0 Ozone
Reagent. Replace the cap and
shake gently for 20 seconds.
Replace the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
measurement. Alternatively, wait for 2 minutes and press READ
DIRECT. In both cases “SIP” will blink during measurement.
The instrument directly displays concentration in mg/L of ozone
on the Liquid Crystal Display (chlorine free samples only).
10 mL