Welch Allyn Medical Diagnostic Equipment 622NP Blower User Manual


 
20
Atlas Monitor showing Impedance Respiration waveform display,
controls, and ECG connector
Setting alarm limits –To set the Respiration
alarm limits, use the
Select and Set buttons on
the left side of the monitor to follow these steps:
Press the left Select button to cycle through
MAP
HI and LO, Heart Rate HI and LO,
Respiration
HI and LO, then (model 623xx
only) CO
2
HI and LO.
Each push of Select moves you to the next
limit. Stop at the limit you want to set. The
measurement and
HI or LO will flash,
indicating which limit is being changed.
Press the Set button up or down to raise or
lower the alarm limit. (When you come to
the end of the range, the numbers will stop
changing).
Press the Select button again to go to the
next limit, or press it several times until
none of the measurements flash and no
HI or
LO indicators are lit. The instrument is now
in its normal measurement mode. (If you do
not press any button for 10 seconds, the
instrument will automatically revert to its
normal measurement mode).