Teledyne 701 Portable Generator User Manual


 
Teledyne API Model 701 Zero Air Generator
A Primer on Electro-Static Discharge
01671H (DCN6051) 8-7
2. Never carry the component or assembly without placing it in an anti-ESD bag or bin.
3. Before using the bag or container allow any surface charges on it to dissipate:
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at an anti-ESD workbench, lay the container on the conductive work surface.
In either case wait several seconds.
4. Place the item in the container.
5. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD tape.
Folding the open end over isolates the component(s) inside from the effects of static
fields.
Leaving the bag open or simply stapling it shut without folding it closed prevents the bag
from forming a complete protective envelope around the device.
6. Once you have arrived at your destination, allow any surface charges that may have built
up on the bag or bin during travel to dissipate:
Connect your wrist strap to ground.
If you are at the instrument rack, hold the bag in one hand while your wrist strap is
connected to a ground point.
If you are at a anti-ESD work bench, lay the container on the conductive work surface
In either case wait several seconds
7. Open the container.
8.5.4 Opening Shipments from Teledyne API’s Customer Service
Packing materials such as bubble pack and Styrofoam pellets are extremely efficient
generators of static electric charges. To prevent damage from ESD, Teledyne API
ships all electronic components and assemblies in properly sealed anti-ESD containers.
Static charges will build up on the outer surface of the anti-ESD container during
shipping as the packing materials vibrate and rub against each other. To prevent these
static charges from damaging the components or assemblies being shipped ensure that
you always unpack shipments from Teledyne API’s Customer Service by:
1. Opening the outer shipping box away from the anti-ESD work area.
2. Carry the still sealed ant-ESD bag, tube or bin to the anti-ESD work area.
3. Follow steps 6 and 7 of Section 8.5.3 above when opening the anti-ESD container at the
work station.
4.
Reserve the anti-ESD container or bag to use when packing electronic components or
assemblies to be returned to Teledyne API.