Teledyne 701 Portable Generator User Manual


 
Teledyne API Model M701 Zero Air Generator
OPTIONAL HARDWARE and SOFTWARE
01671H (DCN6051) 4-3
4.4 Rack Mount Kits (OPT 20A, OPT 20B & OPT 21)
There are several options for mounting the analyzer in standard 19” racks. The slides
are three-part extensions, one mounts to the rack, one mounts to the analyzer chassis
and the middle part remains on the rack slide when the analyzer is taken out. The
analyzer locks into place when fully extended and cannot be pulled out without pushing
two buttons, one on each side.
The rack mount brackets for the analyzer require that you have a support structure in
your rack to support the weight of the analyzer. The brackets cannot carry the full
weight of an analyzer and are meant only to fix the analyzer to the front of a rack,
preventing it from sliding out of the rack accidentally.
OPTION NUMBER DESCRIPTION
20A Rack mount brackets with 26 in. chassis slides.
20B Rack mount brackets with 24 in. chassis slides.
21 Rack mount brackets only.
4.5 Expendable Kits (OPT 42A)
A one-year supply of replacement particulate filters is available from Teledyne API as
Option 42A. Refer to Appendix B for a list of expendables.
4.6 Flow Restriction (13 SLPM) (OPT 72B)
A flow restrictor is available in order to ensure a source of dry air. It is connected to the
outlet fitting and restricts the air flow to 13 SLPM.
4.7 Dewpoint Sensor (OPT 86E)
The dewpoint sensor ensures that the Regenerative Dryer maintains a dewpoint less
than -20°C (up to 15 SLPM flow rate) or less than 10°C (greater than 15 SLPM flow
rate). This will increase the life of the chemical scrubbers.