89
C4 Manual Rev 7.5.2
9. APPENDIX
9.1 Programming Examples & Notes
9.1.1 Simple Local Program Example
The following temperature test routine uses basic temperature control commands
including controlled and maximum rate ramping and temperature hold (dwell):
Using Probe 1 for control...
Use 5 minutes to ramp from ambient to 45.2
/C
Hold 45.2 for 2 minutes
Use 2 minutes to ramp to 32.3
/C
Hold 32.3/C for 6 minutes
Go to 50.1/C as quickly as possible
Go to -10/C as quickly as possible
Hold -10/C for 3 minutes
End program
The following is the exact key sequence for entering this routine as a program into
C4 program memory:
Set the C4 to Program mode by setting the rotary MODE switch to the PROGRAM
position.
If you wish to erase any programs already in memory and start with “clear”
reinitialized program memory, press CLEAR/PROG CLEAR/PROG
The table which follows shows how to enter all of the data for this program. There
are shortcuts that can be used to save some time when entering program data.
The shortcuts are shown in the next table. If you are not familiar with C4
programming, be sure your understand all of the entries in the first table before
attempting to use the shortcut method.