Princeton XS-432 Landscape Lighting User Manual


 
5
Installation of the Xenon Lamp in the Lamp Housing:
(Note: The lamp housing must be mounted to the entrance slit as described above)
-----------------------------------------------------------------------------------------------------
CAUTION
Never touch the xenon lamp (bulb) or the surface of the focusing mirror.
If the xenon lamp should become contaminated, refer to the manufacturer’s
handling precautions for proper cleaning procedure.
-----------------------------------------------------------------------------------------------------
To initially install the xenon lamp in the lamp housing, use the following procedure.
B1. Remove the four (4) screws from the lamp housing cover and remove the
cover.
B2. Locate the 1/16 and 7/64 Allen wrenches supplied.
B3. A shipping post is installed in place of the lamp for shipping purposes.
Inside the lamp housing, locate and become familiar with the shipping post,
the top terminal, the mirror locations and the base terminal with a split
clamp arrangement securing the shipping post.
Caution: Do not touch, talk or breathe on the mirrors.
B4. Note the orientation of the top terminal, as the lamp and top terminal must be
installed in the same orientation. Using the 1/16” Allen wrench, loosen the
set screw in the end of the top. Note: Do not remove the wire from the top
terminal. Remove the top terminal from the shipping post and set the
terminal inside the lamp housing, adjacent to the mirror.
B5. Insert the end of the 7/64” Allen wrench into the screw in the split clamp of
the base terminal and loosen the screw. Remove the shipping post and save
for future use.