General
1.3 Conditions for Use
The use of the I-Transia ceiling hoist is subject to the following:
• The I-Transia should only be used by trained personnel.
• The safe working load (SWL), 205 kg (450 lbs) or 250 kg (550 lbs), must not be
exceeded. Refer to 2.3 Load Limits, page 8
.
• The instruction offered by Invacare to all customer groups in connection with the
purchase of a ceiling-mounted hoist has been received.
• The caregiver pays attention to the well-being of the user when using the hoist.
• The hoist is used in rail systems which are installed, tested and approved according to
DS/EN 10535 and Invacare’s stipulations.
• Only technicians who have been certified by Invacare may install and test the rail systems.
• The hoist is used with an Invacare hanger bar.
• The hoist is used with an Invacare lifting sling or with other suitable slings. Refer to
5.1 Sling Information, page 18
.
1.4 Service Life
The I-Transia has an expected lifetime of 15 years, on the condition of correct use and
correct service inspections. Refer to the I-Transia user manual for more information.
1.5 Features
Information panel on the I-Transia bottom surface
A
Type and maximum load designation
B
Green/yellow battery indicator lamp
C
Strap for activating emergency stop and
emergency lowering device
D
Button for resetting the emergency stop
1183377-B 5