SECTION 4—MAINTENANCE
Part No 1150695 31 Recessed Side Entry Bath Tubs
Heavy Duty Cleaning Procedure
ƽ WARNING
The disinfectant dispensing system has been factory tested and calibrated for use with Invacare
disinfectant. The use of any disinfectant with a dilution ratio (½ oz. per gallon) or viscosity level different
than that of Invacare disinfectant will result in an improper mixture of disinfectant and water. This may
cause your system to be improperly disinfected - or in some instances, not disinfected at all. If using any
disinfectant with a dilution ratio or viscosity level different than that of Invacare disinfectant or if
switching from one type of disinfectant to another type, the disinfectant flowmeter must be calibrated
before performing this procedure. Refer to Calibrating the Disinfectant Flow Meter
on page 32.
ALWAYS wear rubber gloves, an apron and face shield protection during disinfecting, cleaning and
rinsing procedures. In case of eye or skin contact, follow procedures on disinfectant container.
CAUTION
NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The
use of these types of cleaners will make the gel-coat finish of your tub scratched and dull. NEVER use
cleaners containing iodine, bromine, Betadine or methylene blue. The use of these types of cleaners
will cause the tub to stain. Never use cleaners containing bleach. Use of these types of cleaners will
dry out the rubber seals and gaskets and the tub will not function properly.
Before starting the whirlpool pump, the water level must be completely above the whirlpool jets.
Otherwise damage to the motor/pump will result.
Inadditiontodisinfectingyourtubaftereveryuse,youwillneedtoperformthisprocedureapprox imatelyonce
every80150baths.
1. Clean theentiresurfaceofthetubtoremovevisibletissue
residueandfluid.
2. Rinse theinsidesurfaceofthetubwiththeshower/massage
wandonthespraysetting.
3. Rotate
thewatertemperaturevalvetoadjustthewater
temperaturetoapproximately105
o
F.
NOTE:Thiswillmaximizethecleaningsolutionʹsabilitytocleanand
disinfectthesystem.
4. Close thedrain.
5. Make surethewhirlpooljetsareopen.
6. Fill thetuboneinchabovethehighestwhirlpooljets.
7. Add theInvacareHeavyDutyCleanertothewaterinthe
tub.Refertothechartfor
theproperamount.
8. Spray premixedInv acareHeavyDutyCleaneronthe
remainingsurfaces.
9. Press thewhirlpoolon/offbuttontoturnonthewhirlpool
pumpandletthepumprunfor60seconds.
10. TurnoffthepumpandallowtheHeavyDutyCleaner/
watersolutiontoremaininthetubfor10
minutes.
11. Du ringthistenminutewaitingtime,takealonghandled
brushandthoroughlyscruballinteriorsurfacesofthetub.
12. Openthedraintoallow theHeavyDutyCleaner /w ater
solutiontodrainfromthetub.
13. Use theshower/massagewandonthepulsesetting and
directthe sprayintobothwhirlpool
jetstobackflushunt i l
thewaterdis chargi ngfromthewhirlpoolinletis clear.
14. Rinsethetubthoroughlywiththeshow er wandonthespraysetting.
15. Checkthetubfordamageandtoensureallsurfacesofthetubarecleanandthatnovisibleresidue
remains.Ifresidueisfound,repeat
STEPS13and14
TUB
MODEL
NUMBER
WATER REQUIRED
TO COVER JETS
(GALLONS)
INVACARE
HEAVY DUTY
CLEANER
REQUIRED (OZ)*
2800 10 40
2001 32 128
3700 30 120
3750 30 120
3800 30 120
3600 30 120
3650 48 192
3600XL 48 192
2300 25 100
3300 30 120
4300 35 140
4300HL 35 140
5300 52 208
5300HL 52 208
6300 50 200
6300ADH 50 200
4400SS 48 192
5400SS 50 200
*NOTE:IfusingnonInvacareheavyduty
cleaner,refertotheinstructionssuppliedwiththe
cleanertodeterminetheproperamountrequired.