6 Inspection and Troubleshooting
If any part of the system is not functioning properly, DO NOT use the system. For parts and technical service information please contact ICCI
Technical Service at (800) 668-2337, Extension 2.
6.1 Safety Inspection Checklists
Initially
Ensure that the maximum water supply temperature does not exceed 110°F.
Ensure that the maximum water pressure does not exceed 55 PSI.
Verify the water flow temperature with a separate thermometer.
Visually inspect for cracks in the tub surface.
Ensure that the door handle locks the door closed.
Ensure that the safety pin properly engages the door.
Check that all labels are present and legible. Replace if necessary.
Before Each Use
Clean and Disinfect the tub. Refer to Cleaning and Disinfecting on page 23.
Ensure that the maximum water fill temperature does not exceed 110°F (43° C).
Ensure that the maximum water pressure does not exceed 55 PSI.
Visually inspect for cracks in the tub surface.
Ensure that the door handle locks the door closed.
Ensure that the safety pin properly engages the door.
Every Six Months
Verify the water flow temperature with a separate thermometer.
Check that all labels are present and legible. Replace if necessary.
Clean the Water Temperature Control Valve to remove lime deposits.
Every six months or as necessary have a qualified technician perform a thorough inspection and servicing. Routine
maintenance will extend the life and efficiency of your tub.