46 Model IFT9701 Transmitter Instruction Manual
Configuration with ProLink II Software continued
7.2 Configuration parameters
Use the Configuration window to perform the following tasks:
• Assigning a HART tag to the flowmeter
• Changing process variable measurement:
- Changing measurement units for the mass flow or volume flow rate
- Changing the flow cutoff value
- Changing the internal damping value
- Changing the flow direction parameter
• Setting range values for the milliamp output
• Scaling the pulse output
7.2.1 HART tag
The HART tag consists of up to eight characters that identify the flowmeter when it communicates
with other devices in a HART multidrop network.
To assign a HART tag to a flowmeter:
1. From the ProLink menu, click on Configuration.
2. In the Configuration window, click on the Device tab.
3. When the Device panel appears (as depicted in Figure 7-1), enter the desired tag of up to eight
characters, then click on OK. The tag can include spaces and periods.
Figure 7-1 Configuration window – Device panel