Welch Allyn Medical Diagnostic Equipment 622NO Blower User Manual


 
66
A.3 NIBP
Systolic Range 60 to 250 mmHg (8 to 33 kPa)
Diastolic Range 30 to 160 mmHg (4 to 21 kPa)
Mean Arterial Pressure (MAP) Range 40 to 190 mmHg
Blood Pressure Accuracy Per AAMI SP10-1992 Clause 4.4.2
Efficacy study results using Auscultatory Method
as Reference Standard are available upon request.
Heart Rate Range for Specified Accuracy 40 to 200 beats/minute
Display Numeric displays for Systolic, Diastolic and MAP
Cuff pressure displayed during measurement cycle
Data available in Trend display and printed report
Method Oscillometric
Control Manual and automatic control
Automatic Intervals 1, 3, 5, 10, 15, 30, and 60 minutes
Alarms High and Low Systolic
High and Low Diastolic
High and Low MAP (in mmHg)
Alarm Ranges Systolic Low Limits: 60 – 245 mmHg
Systolic High Limits: 65 – 250 mmHg
Diastolic Low Limits: 30 – 150 mmHg
Diastolic High Limits: 35 – 155 mmHg
MAP Low Limits: 45 – 185 mmHg
MAP High Limits 50 – 190 mmHg
Initial Cuff Pressure 120, 140, 160, 180, 200, 240, 280 mmHg - user
selectable
Repeated Cuff Pressure Previous Systolic pressure + 40 mmHg
Maximum Inflation Time 3 minutes (typical: <40 seconds)
Inflation Over-Pressure Dump <300 mmHg
Inflation Over-Pressure Dump Backup <330 mmHg
A.4 ECG
Heart Rate Range 21 to 249 beats/minute
Heart Rate Accuracy
±3 beats/minute or ±3% whichever is greater
Bandwidth, Normal Mode 0.5 to 40 Hz
Bandwidth, Extended Mode 0.05 to 100 Hz
Leads 3 wire or 5 wire, available in AHA or IEC colors
Connector AAMI 6 pin
Electrodes Disposable snap electrodes
Display Sweep Speeds 6.25, 12.5, 25 mm/second
Gain 10 mm/mV, and automatic gain with visual scale
indicator