Invacare I-Transia 250 kg Patio Heater User Manual


 
I-Transia™
Daily Inspections
The hoist should be checked each time it is used. Perform the following checks in addition to
those performed in the annual service inspection. If you question the safety of any part of the
lift, DO NOT use. Contact your Invacare dealer IMMEDIATELY.
q Visually inspect the sling, hanger bar and lifting strap. Check all parts for damage or
wear. If damage is found, DO NOT use. Contact your Invacare dealer IMMEDIATELY.
q Check the emergency stop and emergency lowering function. Check all parts for
external damage or wear. If damage is found, DO NOT use. Contact your Invacare
dealer IMMEDIATELY.
q Check that all hardware and attachment points for damage or wear. Check all parts for
external damage or wear. If damage is found, DO NOT use. Contact your Invacare
dealer IMMEDIATELY.
q Verify that the hand control is functional.
q Check the battery charger.
q Charge the battery every day the lift is used.
q Check the carabiner for wear at the bottom contact point A. Do not use the hoist if the
carabiner measures less than 6 mm (1/4 in) at this point.
Safety Inspection Checklist
The annual safety inspection checklist is available in the I-Transia service manual.
8.3 Cleaning the Sling and Hoist
Cleaning the Sling
Refer to the washing instructions on the sling and to the sling manual for cleaning details.
Cleaning and Disinfecting the Hoist
To prevent cross-infection, the parts of the hoist that come in contact with caregivers and
patients must be cleaned and disinfected.
A soft cloth, dampened with water and a small amount of dish liquid, is all that is needed
to clean the hoist, lifting strap and carabiner. The hoist can be cleaned with non-abrasive
cleaners.
Never use acids, alkaline or solvents for cleaning the hoist. Dry the hoist carefully after
cleaning.
The hoist, hand control and mounting parts can be damaged if the hoist is cleaned any other
way than stated above.
The hoist must be wiped with a moistened, firmly wrung cloth with ordinary household
disinfectants. Only use disinfection detergents approved by the facility.
24