Fluke 271 Portable Generator User Manual


 
271
Getting Started Manual
Front Panel Inputs and Outputs
MAIN OUT
MAIN OUT is the 50 or
600 output from the main
generator. It will provide up
to 20 V p-p into a high-
impedance load or 10 V p-p
into a matched 50 or
600 load. It can tolerate a
short circuit for 60 seconds.
shx0005f.gif
AUX OUT
The AUX OUT socket
provides a two-level TTL
and CMOS compatible
signal, synchronous with
the signal from the
MAIN OUT socket.
shx0006f.gif
The symmetry of the AUX OUT signal is the same as that
set for the main output but you can adjust the phase
relationship between
MAIN OUT and AUX OUT using the
PHASE setting in the TRIG menu.
AUX OUT logic levels are nominally 0 V and 5 V from
typically 50 .
AUX OUT will withstand a short-circuit.
EXT TRIG
EXT TRIG is the external
trigger input for the non-
continuous modes (trigger,
gate, sweep, FSK and hop)
and for synchronizing the
generator as a slave to an
external master generator.
shx0012f.gif
8