Agilent Technologies N5183A MXG Portable Generator User Manual


 
Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide 171
Programming the Status Register System
Status Groups
Data Questionable Frequency Status Group
NOTE Some bits in this status group do not apply to the N5181A/82A/83A, E4428C, E8257D, and
the E8663B and returns zero when queried. See Table 4-9 on page 172 for more information.
The Data Questionable Frequency Status Group is used to determine the specific event that set bit 5
in the Data Questionable Condition Register. This group consists of the Data Questionable Frequency
Condition Register, the Data Questionable Frequency Transition Filters (negative and positive), the
Data Questionable Frequency Event Register, and the Data Questionable Frequency Event Enable
Register.
This is the named
status register for
the E4438C.
However, not all
signal generator
models use all of
the shown events
(i.e. some use only
a subset of the
E4438C’s status
registers).