Agilent Technologies N5183A Portable Generator User Manual


 
Agilent N518xA, E8663B, E44x8C, and E82x7D Signal Generators Programming Guide 343
Creating and Downloading User-Data Files
Data Transfer Troubleshooting (N5182A, E4438C and E8267D Only)
User FIR Filter Coefficient File Download Problems
Data Requirements
Data must be in ASCII format.
Downloads must be in list format.
Filters containing more symbols than the hardware allows (64 for real-time and 512 for Arb) will
not be selectable for the configuration.
Table 8-15 User FIR File Download Trouble - Symptoms and Causes
Symptom Possible Cause
ERROR -321, Out of memory
There is not enough memory available for the FIR coefficient file being
downloaded.
To solve the problem, either reduce the file size of the FIR file or delete
unnecessary files from memory.
ERROR -223, Too much data
User FIR filter has too many symbols.
Real-Time cannot use a filter that has more than 64 symbols (512
symbols maximum for Arb). You may have specified an incorrect
oversample ratio in the filter table editor.