Agilent Technologies 6813B Yard Vacuum User Manual


 
3
Verification and Calibration for Option 019
There are no changes to the output verification and calibration sections of the User's Guide for units with
Option 019.
Option 019 Factory Modifications
Make the following changes to the Agilent 6813B Service Manual (p/n 5962-0859) for Agilent 6813B
Option 019 units.
Initialization
For ROM revision A.00.07, change step 3 to “Scroll to 6813J02 and press Enter”.
NOTE: Option J02 and Option 019 are identical; the J02 option was renamed option 019. After
an Agilent 6813B Option 019 Rev A.00.07 has been re-initialized, the number <1002>
will be returned by a *OPT? query.
For ROM revision A.00.08 or higher, change step 3 to “Scroll to 6813B19 and press Enter”.
NOTE: An Agilent 6813B Option 019 Rev A.00.08 will return <019> for a *OPT? Query.
Replaceable Parts
Add or change the following part numbers in the parts list:
Reference Desig. Part Number Description
A4 5064-0081 AC Input Assembly Tested
A4 C1060 - C1067 0180-4963 Capacitor 3300uF 200V
A9 5064-0083 Output/Relay/Filter Assembly Tested
A9 R552 – R553 0698-3156 Resistor 14.7K 1%
A10 5064-0079 Control Assembly Tested
5964-8149 Appendix Option 019
9320-5540 Label-Information