3M M5560 Landscape Lighting User Manual


 
Troubleshooting:
Condition Remedy
Main indicator light inoperative 1. Check electrical supply.
2. Send unit to an authorized service center.
Curing bulb inoperative 1. Replace bulb.
2. Send unit to an authorized service center.
No time signal tone 1. Send unit to an authorized service center.
Reduced curing performance 1. Check intensity with hand held radiometer.
2. Clean light guide exit tip. Consult
cleaning instructions.
3. Clean optical filter in nose cone with a soft cotton swab.
4. Check intensity output.
5. Check optical filter in nose cone for chips or cracks. If
chips or cracks are present on optical filter, send unit to
an authorized service center.
6. Replace light guide.
7. Replace bulb.
8. Check intensity output.
9. Send unit to an authorized service center.
Handpiece turns on, immediately hear 1. Fan possibly blocked, therefore send unit
5 beeps and handpiece shuts off to an authorized service center.
Unit overheats. Five consecutive beeps, 1. Let unit cool down for 10 minutes.
bulb shuts off during curing cycle Restart bulb. If, after a few minutes the bulb shuts off
again after 5-beep signal, or if the bulb will not restart,
send the unit to an authorized service center.
Curing Performance Test
Lay a clean polyester matrix band on white paper and place a clear or white plastic form on the band.
The form should be a cylindrical hole 3-6 mm diameter through material 2.0-2.5 mm thick. Pack
restorative material into the form, cover with polyester and press flat. Direct the curing light at the top
surface (from less than 2 mm) of the sample and cure for the recommended time. Remove the polyester
strips and examine the condition of cure. The bottom surface should appear glassy and resist scratching
by a sharp explorer, indicating adequate cure. (Consult manufacturer’s instructions for other criteria for
correct cure).
If the surface is soft or dull, repeat the test procedure using incrementally longer curing times until
adequate cure is attained, and adjust curing time accordingly. If the adjusted time exceeds 200% of
normal, verify that the restorative material has not degenerated, or correct curing light problems as
directed in the Troubleshooting table.
Repair
Obtain qualified repair service only from 3M ESPE authorized service facilities.
Contact:
3M Health Care Service Center
Suite 200 Bldg. 502
3350 Granada Ave.North
Oakdale, MN 55128
(800) 292-6298
Replacement Parts and Accessories - Can be purchased from authorized dealer
Item 3M ESPE Stock Number
Bulb/12V, 75W 78-8131-1662-7
Light guide - 8 mm diameter 78-8131-1657-7
Light guide - 13 mm diameter 78-8131-1658-5
Mounting Kit 78-8131-1663-5
Eyeshield - 8 mm 78-8131-1660-1
Eyeshield - 13mm 78-8131-1661-9
Specifications
Note: This manual is used for all versions of the 3M ESPE Elipar 2500, M5560.
Operation of equipment - Continuous Duty
- Unit will operate continuously for 5 minutes before automatic shutoff, at which time
the unit may be restarted immediately
7