LF410/LF600 LF410/LF602
2
Fluid conductivity: 5µS/cm minimum
Fluid temperature:
–10 to +180°C : Ceramic type
(14 to 356 °F)
Note : 120°C (248°F) above is separate type
–10 to +120°C : tefolon PFA
(14 to 248°F)
Ambient temperature:
Without cFMus Approval:
-
20 to + 60 °C (
-
4 to 140 °F)
With cFMus Approval:
-10 to +45°C (14 to 113 °F)
Structure: NEMA 4X (IP 67) Watertight
Power consumption:
17W(27VA) or less
19W(29VA) or less (with PROFIBUS)
Conformance to European Community Directives:
EMC directive 89/336/EEC
The low voltage 93/68/EEC
PED 97/23/EC (Note 1)
Note : See table 2 for detail.
Approved hazardous location certifications:
Model: LF414/LF600 and LF414/LF602
cFMus explosion proof:
FM Class I, Division 2, Groups A,B,C, and D.
FM Class II, Division 2, Groups E, F and G.
FM Class III.
Detector and converter combination:
LF410/LF600: Combined type for standard
specification.
LF410/LF602: Separate type for standard
specification.
LF414/LF600: Combined type with Ex approval
of Class I, Division 2 (cFMus).
LF414/LF602: Separate type with Ex approval
of Class I, Division 2 (cFMus).
Model LF410 and LF414 Detectors
Mounting style: Wafer type
Fluid pressure:
-
0.1 to 2.0MPa
(
-
15 to 300psi, or
-
1.0 to 20 bar)
Note: The test pressure before shipping from the factory
is equal to twice the nominal pressure rating of the
customer specified flange connection during 15
minutes.
Connection flange standards: ANSI 150, ANSI
300, BS10 and 16, DIN PN10 and PN16, JIS10K,
JIS16K and JIS20K
Principal materials:
Case — 25 to 100mm (1” to 4”): stainless steel 15,
150, and 200mm (1/2”, 6”, and 8”): carbon
steel
Linings — 15 to 100mm (1/2” to 4”): Ceramic
tube (std.) & teflon PFA (opt.)
150 and 200mm (6” to 8”): teflon PFA
Electrodes — 316L stainless steel (std.)
Grounding rings — 316 stainless steel (std.)
Note: See Table 3 for optional materials and other
related information.
Measuring tube material — 304 stainless steel
(in case of Teflon PFA lining)
Coating — 25 to 100mm (1” to 4”): no coating
(stainless steel body).
15, 150, and 200mm (1/2”, 6”, and 8”):
phthalic acid resin coating with pearl-gray
colored.
Heat shock resistance : for a ceramic tube detector
Heating: ΔT ≤ 150
°C/0.5sec (302°F/0.5sec)
Cooling: ΔT ≤ 100
°C/0.5sec (212°F/0.5sec)
Note: The above means that the ceramic tube detector
withstands the shock of sudden heating
(temperature difference 150°C or less per
0.5seconds) and sudden cooling (temperature
difference 100°C or less per 0.5seconds).
Dimensions and weights: See Figures 3 to 8.
Cable connection port: for separate type detectors.
Cable glands —
LF410: without cFMus approval
Provided as standard
R(PT) 1/2 male screws.
LF414: with cFMus approval
Not provided
3/4-14NPT male screws are required.
Applicable diameter — 11 to 13mm
(0.433 to 0.512 inch)
Model LF600 and LF602 converters
Input signals
Analog signal — the voltage signal from detector,
proportional to process flow rate (for LF602
separate type converter).
Digital input DI (opt.)
Signal type: 20 to 30Vdc voltage signal
Input resistance: 2.7kΩ
Number of inputs: one point
DI function — One of the following functions can
be assigned to the optional DI signal.
Range switching — Selects either the higher or
lower range in the unidirectional or
bidirectional 2-range setting.
Totalizer control — Starts and stops the built-in