Remote Sensor Bars:
2 x 15pin D-sub jacks are located on the side of the unit for plugging in 1 or
2 additional sensor bars by using remote cables up to 200 feet (61m). The
top sensor bar can also be remote with a cable.
Power Options:
9V alkaline, 7.2V NiMH rechargeable pack (charged in the unit), or AC
adapter wall power cube (AC adaptor will operate the unit or recharge the
NiMH battery pack)
Battery Life:
9V alkaline: 140 hours
Rechargeable Nickel Metal Hydride: 300 hours
(Adding additional sensor bars to the QUESTemp°36 reduces battery life.)
Charge Time (NiMH Battery Pack):
16 hours (charge in the unit)
Safety Approvals:
ETL, cETL: Class I,II,III Groups A,B,C,D,E,F,G, Temperature code T3
DEMKO: EEX ia IIC, T3
CE mark
AIR PROBE ACCESSORY
Range:
0 - 20 meters per second. 0.1m/s increments
Sensor:
Omni directional heated thermistor
Accuracy:
+/- (0.1 m/s + 4%) of measurement value
Battery Life:
6 - 8 hours for fully charged NiMH battery
Charge Time:
9 hours
22
16. APPENDIX A : HEAT EXPOSURE TABLES
ACGIH
Screening Criteria for Heat Stress Exposure. WBGT values in °C (°F).
Acclimatized
Work Very
Demands
Light Moderate Heavy Heavy
100% Work 29.5 27.5 26.0
(85.1) (81.5) (78.8)
75% Work; 30.5 28.5 27.5
25% Rest (86.9) (83.3) (81.5)
50% Work; 31.5 29.5 28.5 27.5
50% Rest (88.7) (85.1) (83.3) (81.5)
25% Work; 32.5 31.0 30.0 29.5
75% Rest (90.5) (87.8) (86.0) (85.1)
Unacclimatized
Work Very
Demands
Light Moderate Heavy Heavy
100% Work 27.5 25.0 22.5
(81.5) (77.0) (72.5)
75% Work; 29.0 26.5 24.5
25% Rest (84.2) (79.7) (76.1)
50% Work; 30.0 28.0 26.5 25.0
50% Rest (86.0) (82.4) (79.7) (77.0)
25% Work; 31.0 29.0 28.0 26.5
75% Rest (87.8) (84.2) (82.4) (79.7)
from "American Conference of Governmental Industrial Hygienists - Threshold
Limit Values and Biological Exposure Indices for 2001"
Reprinted with permission from ACGIH