14 TROUBLESHOOTING
Invacare® Lynx™ L-3 & Invacare® Lynx™ L-4 Scooters
54 Part No. 1143205
14.2 Service Indicator Error Codes
For the following table, refer to FIGURE6.1 on page 25.
FLASH DESCRIPTION MEANING
1
Battery Low The batteries are running low.
• Recharge the batteries.
2
Low Battery Fault The batteries have run out of charge.
• Recharge the batteries.
3
High Battery Fault Battery voltage is too high. This may occur if overcharged &/or traveling down a long slope.
• If traveling down a slope, reduce your speed to minimize the amount of regenerative charging.
• Check the battery and associated connections and wiring.
4
Current Limit Time-out The motor has been exceeding its maximum current rating for too long. This may be due to a faulty motor.
• Check the motor and associated connections and wiring.
• Turn the controller off, leave for a few minutes and turn back on again.
5
Park Brake Fault Either a park brake release switch is active or the park brake is faulty.
• Check the park brake and associated connections and wiring.
• Ensure any associated switches are in their correct positions.
6
Throttle OONAPU The Throttle is out of neutral when turning the controller on.
• Ensure the throttle is in neutral when turning the controller on.
• The Throttle may require re-calibration.
7
Speed Pot Fault The throttle, speed limit pot, SRW or their associated wiring may be faulty.
• Check the throttle and speed pot and associated connections and wiring.
8
Motor Voltage Fault The motor or its associated wiring is faulty.
• Check the motor and associated connections and wiring.
9
Other error The controller may have an internal fault.
• Check all connections & wiring.
10
Push/Rollaway Fault The scooter has exceeded the Push or Rollaway speed limits.
• Turn the controller off and then on again.