Blood/Fluid and Irrigation Warming Systems
Check the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524ACheck the 3M
TM
Ranger
TM
system website to ensure you have the most recent version of this document. www.rangerfluidwarming.com reorder #202524A
CONDITION
CAUSE
SOLUTION
•Unitalerts soon aer plugging it in
(unit does not have to be turned on
for this condition to occur).
•Heatertemperaturerisesto44°Cfor
Model 245 (50°C for Model 247) and
unit shuts down soon aer plugging
it in (unit does not have to be turned
on for this condition to occur).
Over-temperature screw on
bottom of unit is loose or
missing.
Make sure over-temperature screw is
completely tightened. If it is missing,
turn unit o and unplug it. Call 3M
Patient Warming technical support.
CONDITION CAUSE SOLUTION
Alert sounds but unit has been turned
o.
•Independentbackupsafety
system has been activated.
•Over-temperaturescrew
loose.
•Unplugunit.Call3MPatient
Warming technical support.
•Makesureover-temperaturescrewis
completely tightened. If it is missing,
turn unit o and unplug it. Call 3M
Patient Warming technical support.
CONDITION CAUSE SOLUTION
Cannot remove cassette from unit. •Cassetteistoofull,uids
are still being infused, or
clamp is open proximal to
the cassette.
•Warmingunitisbelow
patient level, creating
excessive back pressure.
•Makesureuidisdrainedfrom
cassette before sliding out the
cassette, that uids are no longer
infusing, and that clamp is closed
proximal to the warming cassette.
•Raiseunitabovepatientlevel.
CONDITION CAUSE SOLUTION
Alert indicator light illuminates and
alert sounds, alphanumeric display
alternately ashes a temperature of
33°C or below and the word “LO.”
Under-temperature
condition caused by very
high ow using very cold
uid, or defective heater/
relay.
Alert should stop when temperature
rises above 33°C. If alert continues,
turn o unit, unplug unit and
discontinue use. Call 3M Patient
Warming technical support.
CONDITION CAUSE SOLUTION
Alphanumeric display reads “Er 4” or
“Open.”
Open wire on temperature
sensor.
Do not use unit. Call 3M Patient
Warming technical support.
CONDITION CAUSE SOLUTION
Alphanumeric display reads “Er 5” or
“Open.”
Electrical interference. Remove the unit. Refer to biomedical
technician or call 3M Patient
Warming technical support.