B&K 4071 Portable Generator User Manual


 
BK Precision 4071 User Manual Rev.2.2
16
5.4 Internal FM Mode
Introduction
Internal FM mode generates frequency-modulated signal of fixed amplitude. It used internally generated signal to
modulate frequency of carrier signal. Modulated and modulating frequency, output level, deviation, and DC offset
values could be manually entered from the front panel keypad in this mode. External signal gating options is
available in this mode. Internal FM mode could be entered by pressing “Mode”> “FM” >”1” key at any time.
Internal FM Mode Parameters
The Internal FM mode has the following front panel display:
Figure 5.4-1: Internal FM mode display
1. Modulating Frequency
In this field enter the frequency of the modulating sinusoid. You may enter from 0 Hz to 10,000 Hz in 1 Hz steps.
2. Peak Frequency Deviation
In this field specify the degree to which the modulating signal is allowed to change the carrier frequency. You may
enter from 0 Hz (no change) to 5.0 MHz in 1 Hz steps
This parameter is a peak value. If the deviation were 1 KHz and the carrier frequency were 1 MHz, for example,
then the output frequency will swing between a maximum of 1 MHz + 1 KHz and a minimum of 1 MHz - 1 KHz.
NOTE: If values for the deviation and carrier frequencies are entered such that the output frequency exceeds the 0
to 31.5 MHz range, distortion of the output waveform may result.
3. Carrier Frequency
In this field enter the frequency of the carrier. You may enter from 0 Hz (DC) to 31.5 MHz in 0.01 Hz steps.
NOTE: If values for the deviation and carrier frequencies are entered such that the output frequency exceeds the 0
to 31.5 MHz range, distortion of the output waveform may result.
4. Level
In this field enter the output level, from 4 mVp-p to 5 Vp-p in 1 mV steps or from -44.0 dBm to +18.0 dBm in 0.1
dBm steps.
NOTE: The level specified is a 50O loaded level. This is the level of the signal, which will appear across a 50O
load connected to the SIG Out connector. Into an open circuit, the output swing will be twice the value entered.
Offset
You can enter an offset voltage for the output signal. For more information on output offsets refer to section 4.4.
Ext Gating Input
This TTL compatible input can be used to gate the output signal on or off. A logic high voltage turns off the
output. For further information on the Ext Gating Input, refer to section 2.0.
Int FM: 1,000 Hz
Pk dev: 10,000 Hz
1,000,000.00 Hz -10.0 dBm
1
3
2
4