Invacare 2300 Hot Tub User Manual


 
Heavy Duty Cleaning Procedure
For Invacare Continuing Care Group Tubs
ƽ WARNING
DO NOT use this product or any available optional equipment without first completely reading and understanding these
instructions and any additional instructional material such as owner’s manuals, service manuals or instruction sheets supplied
with this product or optional equipment. If you are unable to understand the warnings, cautions or instructions, contact a
healthcare professional, dealer or technical personnel before attempting to use this equipment - otherwise, injury or damage may
occur.
The disinfectant dispensing system has been factory tested and calibrated for use with Invacare disinfectant. The use of any disinfectant
with a dilution ratio (½ oz. per gallon) or viscosity level different than that of Invacare disinfectant will result in an improper mixture of
disinfectant and water. This may cause your system to be improperly disinfected - or in some instances, not disinfected at all. If using
any disinfectant with a dilution ratio or viscosity level different than that of Invacare disinfectant or if switching from one type of
disinfectant to another type, the disinfectant flowmeter must be calibrated before performing this procedure. Refer to the tub owner’s
manual or to the calibration instruction sheet, part number 1154272, for the calibration procedure.
ALWAYS wear rubber gloves, an apron and face shield protection during disinfecting, cleaning and rinsing procedures. In case of
eye or skin contact, follow procedures on disinfectant container.
CAUTION
NEVER use abrasive cleaners like scouring powder or liquid cleaners containing pumice stone. The use of these types of cleaners
will make the gel-coat finish of your tub scratched and dull. NEVER use cleaners containing iodine, bromine, Betadine
®
or
methylene blue. The use of these types of cleaners will cause the tub to stain. Never use cleaners containing bleach. Use of these
types of cleaners will dry out the rubber seals and gaskets and the tub will not function properly.
Before starting the whirlpool pump, the water level must be completely above the whirlpool jets. Otherwise damage to the
motor/pump will result.
Heavy Duty Cleaning Procedure
Inadditiontodisi nfecting yo urtubafter everyus e,youwill need to perfor mthisprocedure
approximatelyonceevery80150baths.
1. Cleantheentiresurfaceofthetubtoremo v evisibletissueresidueandfluid.
2. Rinsetheins idesu rface ofthetub with the show er/m assagewandonthespray setting.
3. Rotate
thewatertemperaturevalvetoadjustthewatertemperaturetoapproximately
105
o
F.
NOTE:Thiswillmaxim izethecleaningsolutionʹsabilitytocleanand disinfectthes yst em.
4. Closethedrain.
5. Makesurethewhirlpooljetsareopen.
6. Fillthetuboneinchabov ethehighestwhirlpooljets.
7. AddtheIn vacareHeavyDutyCleaner to thewaterinthe tub.Referto the chart for
the
properamo unt.
8. Spraypremix edInvacareHeavyDutyCleanerontherem ainingsurfaces.
9. Pressthewhi rlpoolon/offbuttonto turnon the whirlpoolpump and letthe pump ru n
for60seconds.
10. Turnoffthepump andallow theheavydutycleaner/watersolutiontoremai ninthetub
for10minu tes.
11. Duringthistenminutewaitingtime,takealongha ndle dbrushandthoroughlyscru b
allinteriorsurfacesofthetub.
12. Openthedraintoallowtheheavyduty/w atersolutiontodrainfromthetub.
13. Usetheshow er/massage wandonthepulsesettinganddirectthesprayinto both
whirlpooljets to
backflushun tilthewaterdisc hargingfromthewhirlpool inle tis cl ear.
14. Rinsethetubthoroughlywiththeshow er wandonthespraysetting.
15. Checkthetubfordamageandtoensureallsurfacesofthetubareclean and thatno
visibleresidueremains.Ifresidueis found,re peatSTEPS
13and14.
Maintenance Instructions
SAVE THESE INSTRUCTIONS
Invacare Continuing Care Group www.invacare-ccg.com
USA
One Invacare Way
Elyria, Ohio USA
44036-2125
Canada
994 Hargrieve Rd.
London Ontario
N6E IP5 Canada
All rights reserved. Trademarks are identified by the symbols ™ and ®. All
trademarks are owned by or licensed to Invacare Corporation unless
otherwise noted.
Betadine is a registered trademark of Purdue Pharma L.P.
© 2008 Invacare Corporation
Part No 1154271
For customer service call: 800-668-2337 Rev A - 04/08
TUB
MODEL
NUMBER
WATER REQUIRED
TO COVER JETS
(GALLONS)
INVACARE
HEAVY DUTY
CLEANER
REQUIRED (OZ)*
2800 10 40
2001 32 128
3700 30 120
3750 30 120
3800 30 120
3600 30 120
3650 48 192
3600XL 48 192
2300 25 100
3300 30 120
4300 35 140
4300HL 35 140
5300 52 208
5300HL 52 208
6300 50 200
6300ADH 50 200
4400SS 48 192
5400SS 50 200
*NOTE:IfusingnonInvacareheavyduty
cleaner,refertotheinstructionssuppliedwiththe
cleanertodeterminetheproperamountrequired.