Anritsu 682XXB Portable Generator User Manual


 
3-5 682XXB/683XXB POWER
LEVELS
Table 3-2 is a listing of the Series 682XXB and 683XXB Synthesized
Signal Generator models and their maximum leveled ouput power lev
-
els. Certain test procedures will refer you to this table for the maxi
-
mum leveled output power level setting of the instrument model being
tested.
3-6 682XXB/683XXB MM
PERFORMANCE 682XXB/683XXB
VERIFICATION POWER LEVELS
68XXXB
Model
Frequency
(GHz)
Max Leveled
Output Power
Max Leveled
Output Power
w/Step Attenuator
68X37B 2.0 – 20.0 GHz +13.0 dBm +11.0 dBm
68X45B 0.5 – 20.0 GHz +13.0 dBm +11 dBm
68X47B 0.01 – 20.0 GHz +13.0 dBm +11.0 dBm
68X53B
2.0 – 20.0 GHz
20.0 – 26.5 GHz
+9.0 dBm
+6.0 dBm
+7.0 dBm
+3.5 dBm
68X55B
0.5 – 2.2 GHz
2.2 – 20.0 GHz
20.0 – 26.5 GHz
+13.0 dBm
+9.0 dBm
+6.0 dBm
+11.0 dBm
+7.0 dBm
+3.5 dBm
68X59B
0.01 – 2.0 GHz
2.0 – 20.0 GHz
20.0 – 26.5 GHz
+13.0 dBm
+9.0 dBm
+6.0 dBm
+11.0 dBm
+7.0 dBm
+3.5 dBm
68X63B
2.0 – 20.0 GHz
20.0 – 40.0 GHz
+9.0 dBm
+6.0 dBm
+7.0 dBm
+3.0 dBm
68X65B
0.5 – 2.2 GHz
2.2 – 20.0 GHz
20.0 – 40.0 GHz
+13.0 dBm
+9.0 dBm
+6.0 dBm
+11.0 dBm
+7.0 dBm
+3.0 dBm
68X69B
0.01 – 2.0 GHz
2.0 – 20.0 GHz
20.0 – 40.0 GHz
+13.0 dBm
+9.0 dBm
+6.0 dBm
+11.0 dBm
+7.0 dBm
+3.0 dBm
68X75B
0.5 – 2.2 GHz
2.2 – 20.0 GHz
20.0 – 40.0 GHz
40.0 – 50.0 GHz
+11.0 dBm
+10.0 dBm
+2.5 dBm
+2.5 dBm
+10.0 dBm
+8.5 dBm
0.0 dBm
–1.0 dBm
68X77B
0.01 – 2.0 GHz
2.0 – 20.0 GHz
20.0 – 40.0 GHz
40.0 – 50.0 GHz
+12.0 dBm
+10.0 dBm
+2.5 dBm
+2.5 dBm
+10.0 dBm
+8.5 dBm
0.0 dBm
–1.0 dBm
68X85B
0.5 – 2.2 GHz
2.2 – 20.0 GHz
20.0 – 40.0 GHz
40.0 – 50.0 GHz
50.0 – 60.0 GHz
+11.0 dBm
+10.0 dBm
+2.5 dBm
+2.0 dBm
+2.0 dBm
+10.0 dBm
+8.5 dBm
0.0 dBm
–1.5 dBm
–2.0 dBm
68X87B
0.01 – 2.0 GHz
2.0 – 20.0 GHz
20.0 – 40.0 GHz
40.0 – 50.0 GHz
50.0 – 60.0 GHz
+12.0 dBm
+10.0 dBm
+2.5 dBm
+2.0 dBm
+2.0 dBm
+10.0 dBm
+8.5 dBm
0.0 dBm
–1.5 dBm
–2.0 dBm
Table 3-2. 682XXB/683XXB Maximum Leveled Output Power (1 of 2)